Installation
Requirements
Operating Environment
The 4155C/4156C and the 41501A/B are specified to operate within the following
environmental conditions:
10 °C to 40 °C (with using flexible disk drive) 5 °C to 40 °C
Temperature:
(without using flexible disk drive)
Humidity:
20 % to 80 % RH (with using flexible disk drive)
15 % to 80 % RH (without using flexible disk drive)
non-condensing and wet bulb temperature 29 °C or below
Altitude:
0 to 2000 m
Storaging/Shipping Environment
The 4155C/4156C and the 41501A/B are specified to store/ship within the
following environmental conditions:
-22 °C to 60 °C
Temperature:
Humidity:
5 % to 90 % RH, non-condensing and wet bulb temperature
39 °C or below
Altitude:
0 to 4600 m
2-6
Agilent 4155C/4156C User's Guide Vol.1, Edition 11