Make a Transmission Measurement
1. Connect the equipment as shown in Figure 1-1. Use a known good cable such as the one
that was supplied with your analyzer.
Figure OPSE UP here.
Figure 1-1. Equipment Setup for Transmission Measurement
2. Press
(see the cautions on page 1-6)
PRESET
5
3. Verify that the data trace falls within
HP 8713C/14C result. he HP 8711C/12C/30A should look similar, but end at 1300 MHz.
Figure OP RANS here.
Figure 1-2. Typical P 8713C/14C Transmission Measurement
SCALE
.1
ENTER
5
5
6
0.5 dB of 0 dB. See Figure 1-2 for a typical
.
5
Performance T ests
1-7