Table 8-11
Table 8-12
Specifications
Functions
Calculation Example of Measurement Accuracy using HRSMU
Current
Integration
Range
Time
0.5 sec
10 pA/
100 pA
1 sec
2 sec
0.1 sec
1 nA
0.5 sec
2 sec
0.1 sec
10 nA
0.5 sec
2 sec
10 sec
Calculation Example of Measurement Accuracy using MPSMU/HPSMU
Current
Integration
Range
Time
1 nA
0.1 sec
0.5 sec
2 sec
10 nA
0.1 sec
0.5 sec
2 sec
10 sec
Accuracy:
±
(% of reading)
Maximum value is applicable when the following conditions are satisfied.
Current compliance is smaller than the current range.
Capacitance of the DUT and measurement path is smaller than the maximum
measurement value.
8-24
Maximum Value
100 fF/1 pF
2 pF/20 pF
76 pF/760 pF
700 pF
4.5 nF
18 nF
7 nF
45 nF
180 nF
940 nF
Maximum Value
700 pF
4.5 nF
18 nF
7 nF
45 nF
180 nF
940 nF
±
(offset value)
Agilent 4155C/4156C User's Guide Vol.1, Edition 11
Meas.
Accuracy
Reso.
± (4.2 % + 70 pF)
5 fF
± (4.3 % + 90 pF)
10 fF
± (4.3 % + 130 fF)
20 fF
± (0.84 % + 160 fF)
10 fF
± (0.85 % + 280 fF)
40 fF
± (0.93 % + 740 fF)
200 fF
± (0.84 % + 200 fF)
10 fF
± (0.85 % + 440 fF)
40 fF
± (0.93 % + 1.4 pF)
200 fF
± (1.3 % + 6.2 pF)
1 pF
Meas.
Accuracy
Reso.
± (0.91 % + 170 fF)
10 fF
± (0.94 % + 340 fF)
40 fF
± (1 % + 1 pF)
200 fF
± (0.91 % + 180 fF)
10 fF
± (0.94 % + 480 fF)
40 fF
± (1 % + 1.6 pF)
200 fF
± (1.6 % + 7.6 pF)
1 pF