To Connect The Vsu/Vmu/Pgu Outputs - Agilent Technologies 4155C User Manual

Precision semiconductor parameter analyzers
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Installation
Mounting Connectors

To Connect the VSU/VMU/PGU Outputs

The following figures show connection examples for VSU/VMU and PGU output.
VSU/VMU connector
Use AWG 24 single-strand insulated wire
(Agilent part number 8150-0447) to connect
to a prober, socket, or DUT.
Common
Signal line
Insulator
BNC
Plate
connector
2-40
PGU connector
Regardless of the output impedance setting,
use a low-noise coaxial cable (Agilent part
number 8120-0102) from the connector to a
prober, socket, or DUT.
If you use an 16440A selector, use a
low-noise coaxial cable (Agilent part number
8121-1191).
to
DUT
BNC
Wire
connector
Agilent 4155C/4156C User's Guide Vol.1, Edition 11
Common
Signal line
Insulator
Coaxial cable
Plate
to
DUT

Advertisement

Table of Contents
loading

This manual is also suitable for:

41501a41501b4156c

Table of Contents