Installation
Installing Accessories
Connecting the 4155C and the 16442A/B test fixture
Connect the 4155C and the 16442A/B as shown below.
To prevent electrical shock during use, be sure to connect the interlock cable.
4. MPSMUs
Force
4155C Terminals
Intlk
Intlk
VSU
VSU
VMU
VMU
MPSMU
SMU
There are potentially hazardous voltages of up to ±100 V at the Force and
Guard terminals. To prevent electrical shock, do not expose these lines.
Never connect the Guard terminal to any output, including circuit common, frame
ground, or any other guard terminal. Doing so will damage the SMU.
2-14
2. VSUs
3. VMUs
1. Intlk
4155C
16442A/B
Terminals
3 m or 1.5 m Interlock Cable
3 m or 1.5 m Coaxial Cable
3 m or 1.5 m Coaxial Cable
3 m or 1.5 m Triaxial Cable
Agilent 4155C/4156C User's Guide Vol.1, Edition 11