Replacing Assemblies

To replace the circuit board

86
1
Remove the logic analyzer cables using the "To remove the
logic analyzer cable" procedure on
Remove the four screws attaching the ground spring and
2
back panel to the circuit board, then remove the back
panel and the ground spring.
Replace the faulty circuit board with a new circuit board.
3
Position the ground spring and back panel on the back
4
edge of the replacement circuit board. Install four screws
to connect the back panel and ground spring to the
circuit board.
Install the logic analyzer cables using the procedure "To
5
install the logic analyzer cable" on page 84.