Table 8-2
Table 8-3
Voltage
Set
Range
Reso.
± 2 V
100 μV
± 20 V
1 mV
± 40 V
2 mV
± 100 V
5 mV
a. Accuracy: ± (% of reading or setting value + offset value).
b. Vout is the output voltage in volts.
Specifications
Hardware
Multiplier Factor of the Offset Current at 10 pA/100 pA Range
Temperature
5 to 60 % RH
5 to 18 °C
18 to 28 °C
28 to 40 °C
HRSMU Voltage Range, Resolution, and Accuracy
a
Set Accuracy
± (0.02 % + 400 μV)
± (0.02 % + 3 mV)
± (0.025 % + 6 mV)
± (0.03 % + 15 mV)
8-6
Humidity
60 to 80 % RH
×
2
×
1
×
2
Measurement
Meas.
Reso.
Accuracy
2 μV
± (0.01 % + 200 μV)
20 μV
± (0.01 % + 1 mV)
40 μV
± (0.015 % + 2 mV)
100 μV
± (0.02 % + 5 mV)
Agilent 4155C/4156C User's Guide Vol.1, Edition 11
×
2
×
2
×
5
Max. Current
a
100 mA
100 mA
100 mA (Vout ≤ 20 V)
50 mA
(20 V < Vout ≤ 40 V)
100 mA (Vout ≤ 20 V)
50 mA
(20 V < Vout ≤ 40 V)
20 mA
(40 V < Vout ≤ 100 V)
b