Installation
Installing Accessories

To Install the 16441A R-Box

Before installing the 16441A R-box, turn off the 4155C/4156C and 41501A/B.
1. Connect the To R-Box terminal on the 4155C/4156C to the Control terminal on
the 16441A rear panel using a 3.0 m or 1.5 m control cable.
2. Connect the instrument measurement terminals to the 16441A input terminals as
follows:
Instrument
terminals
4155C MPSMU
4156C HRSMU
41501A/B MPSMU
41501A/B HPSMU
3. Connect the 16441A output terminals to the 16442A/B input terminals, or to the
connector plate input terminals, using 40 cm triaxial cables as shown below:
16441A Output
a
Output Force
b
Output Force
b
Output Sense
a. For non-Kelvin connection.
b. For the Kelvin connection, two triaxial cables must be used to con-
nect the 16441A and the 16442A/B, instead of a Kelvin triaxial
cable.
2-18
16441A Input
Input Force
Input
(Force/Sense)
Input Force
Input
(Force/Sense)
16442A/B Input or Connector Plate Input
SMU1, 2, 3, 4, 5, or 6
SMU1
SMU2
Agilent 4155C/4156C User's Guide Vol.1, Edition 11
Cable
3 m or 1.5 m Triaxial Cable
3 m or 1.5 m Kelvin Triaxial Cable
3 m or 1.5 m Triaxial Cable
3 m or 1.5 m Kelvin Triaxial Cable
SMU3
SMU5
SMU4
SMU6