Agilent Technologies 4155C User Manual page 224

Precision semiconductor parameter analyzers
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Specifications
Hardware
PGU (Pulse Generator Unit) Specifications
General
Modes
Amplitude
Window
Maximum current
Output impedance 50 Ω and low impedance (
Burst count range
Trigger output
Pulse/dc Output Voltage and Accuracy
See Table 8-8.
Timing Parameters
See Table 8-9.
Accuracy of Timing parameters
The following data is applied when leading time and trailing time are the same
value.
Period
Width
Delay
Leading Time and Trailing Time
See Table 8-10.
Restrictions of Pulse Timing Parameters Setup
Pulse width < Pulse period
Delay time < Pulse period
Leading time < Pulse width × 0.8
Trailing time < (Pulse period - Pulse width) × 0.8
8-16
pulse or constant
0 Vpp to 40 Vpp
-40.0 V to +40.0 V
±
200 mA (width:
±
100 mA
1 to 65535
Level: TTL
Timing: same timing and width as PGU1 pulse output.
±
(2 % + 2 ns)
±
(3 % + 2 ns)
±
(2 % + 40 ns)
Agilent 4155C/4156C User's Guide Vol.1, Edition 11
1 ms, average current
1 Ω )
100 mA)