Agilent Technologies 4155C User Manual page 226

Precision semiconductor parameter analyzers
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Table 8-9
Range
Table 8-10
Specifications
Hardware
Timing Parameters
a
Period
2 μs to 100 μs
1
100 μs to 1000 μs
2
3
1 ms to 10 ms
4
10 ms to 100 ms
5
100 ms to 1000 ms
6
1 s to 10 s
a. Period, width, and delay of PGU1 and PGU2 must be in the same range.
Leading Time and Trailing Time
a
Leading/Trailing Time
Range
1
100 ns to 1000 ns
0.5 μs to 10.0 μs
2
5.0 μs to 100.0 μs
3
50 μs to 1000 μs
4
5
0.5 ms to 10.0 ms
a. Leading time and trailing time for a PGU must be in the same range.
8-18
Width
1 μs to 99.9 μs
1 μs to 999 μs
0.01 ms to 9.99 ms
0.1 ms to 99.9 ms
1 ms to 999 ms
0.01 s to 9.99 s
Agilent 4155C/4156C User's Guide Vol.1, Edition 11
Delay
0 to 100 μs
0 to 1000 μs
0 to 10 ms
0 to 100 ms
0 to 1000 ms
0 to 10 s
Set
Accuracy
Resolution
±
1 ns
(5 % + 10 ns)
±
10 ns
(5 % + 10 ns)
±
100 ns
(5 % + 10 ns)
1 μs
±
(5 % + 10 ns)
10 μs
±
(5 % + 10 ns)
Set
resolution
0.1 μs
1 μs
10 μs
100 μs
1 ms
10 ms

Advertisement

Table of Contents
loading

This manual is also suitable for:

41501a41501b4156c