Table 8-8
PGU supplemental information
Overshoot (Undershoot,
ringing)
Pulse width jitter
Pulse period jitter
Slew rate
Noise
Pulse/dc Output Voltage and Accuracy
Set
Voltage
Parameter
Range
± 20 V
b
Base
± 40 V
± 20 V
Pulse
± 40 V
a. Accuracy is specified at leading = trailing time = 1 μs.
b. dc output is performed by the Base parameter.
Agilent 4155C/4156C User's Guide Vol.1, Edition 11
≤ ±
5 % of amplitude
±
(50 Ω output impedance to 50 Ω load)
0.2 % + 100 ps
0.2 % + 100 ps
maximum 100 V/μs
(50 Ω output impedance to 50 Ω load)
0.2 % of range (at dc output)
Resolu
-tion
±
4 mV
(1 % of Base + 50 mV + 1 % of Pulse)
±
8 mV
(1 % of Base + 50 mV + 1 % of Pulse)
±
4 mV
(3 % of Base + 50 mV)
8 mV
±
(3 % of Base + 50 mV)
Specifications
Hardware
10 mV
a
Accuracy
8-17