Agilent Technologies 4155C User Manual page 79

Precision semiconductor parameter analyzers
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

NOTE
CAUTION
WARNING
Low-Noise Coaxial Cable
For the extended measurement paths over the connector plate, use low-noise coaxial
cable (Agilent part number 8121-1191). This cable can maximize the guard effects
and minimize the impression of the external noise.
Following figure shows the cutting example of this cable. Key point is the isolation
between the conductive layer and the center conductor. So, cut and trim the end of
the cable as shown in this figure by using a cutter and so on.
Insulator (black)
Outer conductor
(for Guard signal)
Conductive layer (black)
Never connect the Guard terminal to any output, including circuit common, chassis
ground, or any other guard terminal. Doing so will damage the SMU.
Potentially hazardous voltages, up to ±100 V (±200 V for HPSMU), are present
at the Force, Sense, and Guard terminals.
To prevent electrical shock, do not expose these lines.
Before turning the 4155C/4156C on, connect the Intlk terminal to a switch that
turns off when the shielding box access door is opened.
Before you touch any connections to these terminals, turn the 4155C/4156C off,
disconnect the power cable, and discharge any capacitors.
Agilent 4155C/4156C User's Guide Vol.1, Edition 11
Cover here using sleeve
min 2 mm
Insulator (clear)
Installation
Mounting Connectors
Center conductor
(for Force/Sense signal)
approx. 10 to 15 mm
2-39