WARNING
Installation
This chapter describes how to install Agilent 4155C/4156C and accessories.
Requirements
Inspection
Connecting the 41501A/B
Installing accessories
Mounting connectors
Connecting peripherals
Maintenance
There are potentially hazardous voltages (200 V for HPSMU, and 100 V for
HRSMU and MPSMU) present at the Force, Guard, and Sense terminals of the
4155C/4156C. To prevent electrical shock, the following safety precautions
must be observed during the use of the 4155C/4156C.
Use a three-conductor ac power cable to connect the cabinet (if used) and
the 4155C/4156C to an electrical ground (safety ground).
If you do not use the 16442A/B test fixture, you must connect the Intlk
terminal to a switch that turns off when the shielding box access door is
opened.
Confirm periodically that the interlock function works normally.
Before touching the connections on the Force, Guard, and Sense
terminals, turn the 4155C/4156C off and discharge any capacitors. If you
do not turn the 4155C/4156C off, complete all of the following items,
regardless of the 4155C/4156C settings.
Terminate measurement by pressing Stop key, confirm that the
MEASUREMENT indicator is not lit.
Deactivate standby mode (if used) by pressing Standby key, confirm
that the Standby indicator is not lit.
Confirm that the HIGH VOLTAGE indicator is not lit.
Open the shielding box access door (open the Intlk terminal).
Discharge any capacitors connected to an SMU.
Warn persons working around the 4155C/4156C about dangerous
conditions.
2-2
Agilent 4155C/4156C User's Guide Vol.1, Edition 11