Connecting the 4156C and the 16442A/B test fixture
Connect the 4156C and the 16442A/B as shown below.
To prevent electrical shock during use, be sure to connect the interlock cable.
4. HRSMUs
Force
Sense
4156C Terminals
Intlk
Intlk
VSU
VSU
VMU
VMU
HRSMU
SMU1 and SMU2
SMU3 and SMU4
SMU5 and SMU6
There are potentially hazardous voltages of up to ±100 V at the Force, Sense,
and Guard terminals. To prevent electrical shock, do not expose these lines.
Never connect the Guard terminal to any output, including circuit common, chassis
ground, or any other guard terminal. Doing so will damage the SMU.
Agilent 4155C/4156C User's Guide Vol.1, Edition 11
2. VSUs
3. VMUs
1. Intlk
4156C
16442A/B
Terminals
3 m or 1.5 m Interlock Cable
3 m or 1.5 m Coaxial Cable
3 m or 1.5 m Coaxial Cable
3 m or 1.5 m Kelvin Triaxial Cable
(for Kelvin connection) or
3 m or 1.5 m Triaxial Cable (for
non-Kelvin connection)
Installation
Installing Accessories
16442A/B
Cable
2-15