•
Connecting the 16440A selector to the 4155C/4156C
Turn off the 4155C/4156C and 41501A/B before connecting the instruments.
Then connect as shown below.
Instrument
41501A/B
To SMU/Pulse
Generator Selector
Interface
PGU
MPSMU or HPSMU
4155C
MPSMU
4156C
HRSMU
16442A/B
SMU
Connector Plate
SMU
a. You can use two inputs for one selector, and four inputs for two selectors.
b. You can use two outputs for one selector and four outputs for two selectors. Selector
output is either one of the PGU outputs or the SMU output.
Agilent 4155C/4156C User's Guide Vol.1, Edition 11
4155/4156
4
2
41501
16440A
Terminal
Terminal
CONTROL Input
Input PGU
Input SMU
Output Selected
Installing Accessories
16442A/B Test Fixture
3
1
3.0 m or 1.5 m Control Cable
3.0 m or 1.5 m Coaxial Cable
a
3.0 m or 1.5 m Triaxial Cable
40 cm Triaxial Cable
b
Installation
16440A Selector
Cable
2-27