11. CAL MXR Adjustment Routine

This adjustment applies to:
8592L spectrum analyzers
8593E spectrum analyzers
8595E spectrum analyzers
8596E spectrum analyzers
The
softkey accesses an internal adjustment routine which
CAL MXR
optimizes the dc bias for the A3A6 Dual Band Mixer when in high band
(2.75 to 22 GHz). The CAL YTF routine must be performed prior to this
adjustment. New frequency response correction constants must be
developed following the CAL MXR Adjustment Routine.
Equipment Required
Cable, Type N (m) to SMA (m)
Additional Equipment for Option 026
Cable, CAL Comb
Adapter, Type APC-3.5 (f) to APC-3.5 (f)
Additional Equipment for Model 8595E
Cable, BNC, 20 cm (9in)
Chapter 2
Making Adjustments
11. CAL MXR Adjustment Routine