A16A1Bt1 Battery - Agilent Technologies 8590 Series Service Manual

Analyzers assembly-level repair
Table of Contents

Advertisement

A16A1BT1 Battery

This procedure applies to:
Use ESD precautions when performing this replacement procedure.
CAUTION
Battery A16A1BT1 contains lithium iodide. There is a danger of
WARNING
explosion if the battery is incorrectly replaced. Replace only
with the same or equivalent type recommended. Do not
incinerate or puncture this battery. Discard used batteries
according to the manufacture's instructions.
Removal/Replacement
The A16A1 memory board assembly may be damaged if it is placed on a
CAUTION
conductive surface. Use a static-safe workstation. Ensure that the
memory board assembly is not placed on any conductive material.
1. Remove the A16A1 memory board assembly by lifting it up from the
Placing the bottom side of the A16A1 assembly on a conductive work
CAUTION
surface can result in loss of correction data stored in RAM. Protect the
A16A1 assembly from contact with any conductive surface.
2. Remove the nylon insulator from the bottom of the A16A1 board
3. Locate the battery leads on the A16A1 assembly, desolder them, and
Chapter 7
8590 E-Series and L-Series spectrum analyzers
8591C cable TV analyzers
8594Q QAM analyzers
A16 processor/video board.
assembly.
remove the old battery.
Replacing Major Assemblies
A16A1BT1 Battery
337