Making Adjustments: If 3335A Source Not Available
6a. Cal Attenuator Error
Calculate Actual Attenuation Errors
1. From the calibration data supplied with the 1 dB step attenuator,
2. Subtract column 1 from column 2 and enter these values in
Table 2a-1
Calculating Actual Attenuation Errors
Nominal 1 dB Step
Attenuator Setting
Figure 2a-1
Cal Attenuator Error Connection Setup
180
enter in column 2 of
corresponding nominal attenuation settings in column 1.
column 3.
Column 1
1 dB
4 dB
5 dB
7 dB
8 dB
9 dB
Table 2a-1
the actual attenuation for the
Actual Attenuation
Column 2
Attenuator Error
(Column 2 - Column 1)
Column 3
Chapter
2a