Making Adjustments
18. Modulator Offset and Gain for Option 010 or 011
13.Press the following analyzer keys to execute the screen title
command.
,
CAL
More 1 of 4
SERVICE CAL
EXECUTE TITLE
Note that TGMOD should appear in the upper-left screen annotation
below REF.
14.Press the following analyzer keys.
,
AMPLITUDE
SCALE LOG LIN
AUX CTRL, TRACK GEN
15.Adjust the front-panel knob for 0.0 V ±0.01 V on the DMM, then
press the following keys.
,
TRACE
CLEAR WRITE A, More 1 of 3, DETECTOR SAMPL PK
MKR, MARKER ∆
16.Adjust A7A1R49 so that the MKR reading drops approximately 0.1
to 0.15 dB.
17.Press the following analyzer keys.
MKR, MARKER ∆
CAL, More 1 of 4
,
AMPLITUDE
SCALE LOG LIN
AUX CTRL, TRACK GEN
18.Adjust the front-panel knob for −1.0 V ±0.01 V measured with the
DMM, then press the following analyzer keys.
TRACE, CLEAR WRITE A
19.Adjust A7A1R43 so that the MARKER DELTA reading is −33 dB
±0.5 dB.
20.Press
, then wait for the preset routine to finish.
PRESET
21.Perform the tracking generator self-cal routine by pressing the
following analyzer keys.
,
CAL
More 1 of 4
Wait for the completion of the routine (less than 1 minute).
22.Press
.
CAL STORE
128
,
More 2 of 4
(LOG), 1,
,
SRC PWR ON OFF
,
MARKER ∆
,
,
More 2 of 4
SERVICE CAL
(LOG) 10
,
SRC PWR ON OFF
,
,
More 1 OF 4
,
,
More 2 of 4
CAL TRK GEN
dB
(ON)
(PK)
,
EXECUTE TITLE
dB
(ON)
DETECTOR SAMPL PK
Chapter 2
(PK)