Agilent Technologies 8590 Series Service Manual page 155

Analyzers assembly-level repair
Table of Contents

Advertisement

Table 2-12
Frequency Response Attenuator 20 dB
Synthesized
Sweeper Frequency
(MHz)
948
881
836
818
Frequency Response Input Attenuator 30 dB
23.On the analyzer, press the following keys.
24.Set the FREQUENCY of the analyzer to the measurement frequency
25.On the analyzer, press
26.On the synthesized sweeper, press CW and set the frequency to the
27.On the synthesized sweeper, press POWER LEVEL and adjust the
28.Set the power sensor cal factor (for frequency being measured) on
29.Repeat
Table 2-13
Frequency Response Attenuator 30 dB
Synthesized
Sweeper Frequency
(MHz)
818
836
881
948
Chapter 2
Min (dBm)
+0.3
+0.3
+0.3
+0.3
AMPLITUDE, ATTEN AUTO MAN
, 10,
AMPLITUDE
+dBm
shown in
Table
2-13.
same measurement frequency as the analyzer is set in the previous
step.
amplitude so the analyzer marker amplitude reads
+5 dBm ±0.05 dB.
the measuring receiver, then record the measuring receiver power
reading in
Table
2-13.
step 24
to
step 28
948 MHz. Record the results in
Min (dBm)
+4.3
+4.3
+4.3
+4.3
25. Checking the Absolute Amplitude Accuracy for Option 050
Measuring Receiver
Reading (dBm)
_________________
_________________
_________________
_________________
, 30, dB
.
PEAK SEARCH
for frequencies of 836 MHz, 881 MHz, and
Table
2-13.
Measuring Receiver
Reading (dBm)
_________________
_________________
_________________
_________________
Making Adjustments
Max (dBm)
+1.7
+1.7
+1.7
+1.7
Max (dBm)
+5.7
+5.7
+5.7
+5.7
155