Agilent Technologies 8590 Series Service Manual page 641

Analyzers assembly-level repair
Table of Contents

Advertisement

Analyzer Messages
Interpreting Analyzer Messages
INVALID <file name> NOT FOUND (U)
Indicates that the specified file could not be loaded into analyzer
memory or purged from memory because the file name cannot be found.
INVALID FILENAME_ _ _(U)
Indicates the specified file name is invalid. A file name is invalid if
there is no file name specified, if the first letter of the file name is not
alphabetic, or if the specified file type does not match the type of file.
See the description SAVRCLW or STOR programming command for
more information.
INVALID FILE: NO ROOM (U)
Indicates that there is insufficient space available on the memory card
to store the data.
INVALID IB ADRS/OPERATION (U)
An IB operation was aborted due to an incorrect address or invalid
operation.
Related Assemblies
A16 processor/video assembly, A20 IB interface board assembly.
Troubleshooting Hints
Check that there is only one controller (the analyzer) connected to the
printer or plotter.
INVALID IB OPERATION REN TRUE (U)
The IB operation is not allowed.
Related Assemblies
A16 processor/video assembly, A20 IB interface board assembly.
Troubleshooting Hints
This is usually caused by trying to print or plot when a controller is on
the interface bus with the analyzer. To use the analyzer print or plot
functions, you must disconnect any other controllers on the IB. If you
are using programming commands to print or plot, you can use an
BASIC command instead of disconnecting the controller. Ensure the
analyzer is configured properly for PRINT or PLOT, B/W or Paintjet,
and the printer or plotter address is set correctly set. See the
8590 E-Series and L-Series Spectrum Analyzers, and 8591C Cable TV
Analyzer Programmer's Guide for more information.