Agilent Technologies 8590 Series Service Manual page 265

Analyzers assembly-level repair
Table of Contents

Advertisement

To check linear gain control lines for the A14 log amplifier
Table 5-5 provides the control line voltages from the A7 analog interface
assembly to the A14 log amplifier assembly. When the analyzer is in
linear mode, the seven A14 log amplifier stages are biased to operate as
linear amplifiers. Four of the seven stages are also used to provide
40 dB of gain in 10 dB increments. The linear gains are enabled for the
reference levels indicated in
stages operate as one 20 dB amplifier and are controlled by the same
control line, IFG6.
Refer to
chapter when tracing control lines in the IF section.
When enabled, each control line has a −7.6 Vdc output. This voltage is
supplied by the −8 VT temperature-compensated power supply located
on the A14 assembly.
The entries in
follows:
Table 5-5
Linear Gain Control Lines on the A14 Assembly
Reference
Level
(dBm)
−50
−60
−70
−80
−90
H = +14.3 Vdc (disabled)
L = −7.6 Vdc (enabled)
Chapter 5
"Troubleshooting the A15 Motherboard Assembly"
Table 5-5
PRESET
SCALE LOG LIN, LIN
,
AMPTD UNITS
dBm
CAL
More 1 of 3
(OFF)
CORRECT ON OFF
A14 Gain in
Linear Mode
(10-1 dB Step)
(dB)
A14P1-38
0
10
20
30
40
Table 5-2
and
Table
are valid when the instrument is set up as
IFG4
IFG5
(10-2 dB Step)
A14P1-39
H
H
L
H
L
L
L
H
L
L
Troubleshooting the IF Section
IF Power-Level Measurement
5-5. Two of the four
in this
IFG6
(20 dB Step)
A14P1-40
H
H
H
L
L
265