For continued protection against fire hazard replace line fuse
WARNING
only with same type and rating (F 5A/250V). The use of other
fuses or material is prohibited.
The power cord is connected to internal capacitors that may
WARNING
remain live for 10 seconds after disconnecting the plug from its
power supply.
The detachable power cord is the instrument disconnect
WARNING
device. It disconnects the mains circuits from the mains supply
before other parts of the instrument. The front panel switch is
only a standby switch and is not a LINE switch (disconnect
device).
Before switching on this instrument, make sure that the line voltage
CAUTION
selector switch is set to the voltage of the power supply and the correct
fuse is installed.
Chapter 16
Safety and Regulatory Information
General Safety Considerations