Agilent Technologies 8590 Series Service Manual page 563

Analyzers assembly-level repair
Table of Contents

Advertisement

BINARY SPAN Binary Span
D
Note that this function is available for 8590L and
ESCRIPTION
8591E spectrum analyzers and 8591C cable TV
analyzers only.
Displays the output of the Span DAC on the A7 analog
interface assembly. A ramp is displayed from the
upper-left corner to the lower right corner of the
display.
The instrument span must be set manually before
enabling
The slope of the ramp increases within each of the
following span ranges.
• 0 Hz to 170 kHz
• >170 kHz to 1.7 MHz
• >1.7 MHz to 10 MHz
• >10 MHz to 170 MHz
• >170 MHz to 1.8 GHz
Refer to
information.
R
S
ELATED
OFTKEYS
CAL Calibration Key
D
CAL
ESCRIPTION
self-calibration, service-diagnostics, and
service-calibration functions. A passcode is required for
access to specific softkeys.
R
S
ELATED
OFTKEYS
Refer to
this chapter for the softkeys that are available after
pressing
Chapter 13
.
BINARY SPAN
Chapter 6
for further troubleshooting
,
MAIN SPAN
FM SPAN
provides access to the softkeys for the
Figure 13-1
and
Figure 13-2
.
CAL
Softkey Descriptions
Softkey Descriptions
at the beginning of
563