Agilent Technologies 8590 Series Service Manual page 239

Analyzers assembly-level repair
Table of Contents

Advertisement

Verify the displayed calibration data
1. Press the following analyzer keys.
CAL, More 1 of 4, More 2 of 4
SERV DIAG
DISPLAY CAL DATA
2. Verify that the digital correction values are within their limits. Refer
to
Chapter 14
for typical examples of the calibration data displays.
• If default values are being used, all digital corrections should be
zero.
• If corrections are being used, make sure that all digital
corrections are present. Typical values should be below 1.00.
3. Return to the diagnostics menu by pressing the following analyzer
keys.
PRESET
FREQUENCY, BAND LOCK, 0-2.9 GHz BAND 0
8591E, and 8594E)
CAL, More 1 of 4, More 2 of 4
SERV DIAG, More 1
Chapter 4
Troubleshooting the Analyzer
Using the Internal Service-Diagnostic Routines
(except 8590L,
239