3. Adjust the position of the horizontal trace to
center-screen using the A1A2 knob.
4. To calculate the actual frequency offset in kHz,
multiply the displayed value by 300.
R
S
ELATED
OFTKEYS
DEMOD AM FM
FM GAIN
FM SPAN FM Span
D
Displays the FM_SPAN signal from the Span Dividers
ESCRIPTION
on the A7 analog interface assembly.
Perform the following steps to observe the FM-SPAN
signal for FM spans (LO spans less than or equal to
10 MHz).
1. Activate
SPAN
2. Press
3. Adjust the span setting while observing the
displayed ramp. The slope of the ramp increases as
the span is increased.
Due to quantization errors, the display appears flat for
LO spans less that 500 kHz.
For the main-coil spans (LO spans greater than
10 MHz) the display is a flat line.
Refer to
information.
R
S
ELATED
OFTKEYS
FREQ DIAG Frequency Diagnostics
D
Displays, in real-time, frequency diagnostic information
ESCRIPTION
for the LO section. Refer to
of each block of frequency data.
Chapter 13
DEMOD
before pressing
SPAN
function.
to display a negative-going ramp.
FM SPAN
Chapter 6
additional LO troubleshooting
MAIN SPAN
Softkey Descriptions
Softkey Descriptions
to select the
CAL
Figure 13-9
for the location
FM