Agilent Technologies 8590 Series Service Manual page 148

Analyzers assembly-level repair
Table of Contents

Advertisement

Making Adjustments
24. Tracking Oscillator for Option 010
Figure 2-37
Tracking Oscillator Adjustment Setup
15.Set the analyzer
SRC PWR ON OFF
minutes. Set the controls as follows:
16.Set the microwave frequency counter controls as follows:
17.Remove the screw, located on the front of the tracking generator,
used to seal the tracking oscillator adjustment.
18.On the analyzer, press
Hz
19.Record the microwave frequency counter reading in
20.On the analyzer, press
21.Record the microwave frequency counter reading in
148
switch to on. Press
LINE
(ON). Allow the analyzer to warm up for at least 5
CENTER FREQ ............................................ 300 MHz
SPAN .....................................................................0 Hz
SAMPLE RATE ......................................... Fully CCW
10 Hz-500 MHz SWITCH ................. 10 Hz-500 MHz
500 MHz-26.5 GHz SWITCH ............ 10 Hz-500 MHz
50 Ω - 1 MΩ SWITCH ..........................................50 Ω
.
AUX CTRL
,
AUX CTRL
TRACK GEN
, 4095,
MAN TRK ADJUST
,
TRACK GEN,
,
, 0,
MAN TRK ADJUST
Table 2-9
as F1.
.
Hz
Table 2-9
as F2.
Chapter 2