Agilent Technologies 8590 Series Service Manual page 169

Analyzers assembly-level repair
Table of Contents

Advertisement

Frequency Response Input Attenuator 30 dB
50.On the analyzer, press the following keys.
51.Set the FREQUENCY of the analyzer to the measurement frequency
52.On the analyzer, press
53.On the synthesized sweeper, press CW and set the frequency to the
54.On the synthesized sweeper, press POWER LEVEL and adjust the
55.Set the power sensor cal factor (for frequency being measured) on
56.Repeat
Table 2-24
Frequency Response Attenuator 30 dB
Synthesized
Sweeper Frequency
(MHz)
1429
1453
1477
1501
Chapter 2
,
AMPLITUDE
ATTEN AUTO MAN
, 10,
AMPLITUDE
+dBm
shown in
Table
2-24.
same measurement frequency as the analyzer is set in the previous
step.
amplitude so the analyzer marker amplitude reads +5 dBm
±0.05 dB.
the measuring receiver, then record the measuring receiver power
reading in
Table
2-24.
step 51
to
step 55
and 1501 MHz. Record the results in
Min (dBm)
+4.4
+4.4
+4.4
+4.4
27. Checking the Absolute Amplitude Accuracy for Option 051
, 30, dB
.
PEAK SEARCH
for frequencies of 1453 MHz, 1477 MHz,
Table
Measuring Receiver
Reading (dBm)
_________________
_________________
_________________
_________________
Making Adjustments
2-24.
Max (dBm)
+5.6
+5.6
+5.6
+5.6
169