Making Adjustments
10. CAL YTF Adjustment Routine
1. Press the following keys to activate title mode.
2. Record the displayed "Optimum Bias" DAC values for each
Table 2-5
Acceptable Mixer Bias DAC Correction Values
Band
1
2
3
4
3. If the recorded "Optimum Bias" values are within the acceptable
4. Press the following keys to allow entry of the default mixer bias DAC
5. Enter CALMXRDATA 1600,1; as a title entry to set the mixer bias
6. Press the following keys to store the mixer bias value for band 1 in
74
PRESET
,
CAL, More 1 of 4
More 2 of 4
SERVICE DIAG
DISPLAY CAL DATA
frequency band.
Optimum Bias
_________________
_________________
_________________
_________________
ranges indicated in
Table
necessary.
If the recorded "Optimum Bias" values are not within the acceptable
ranges indicated in
Table
initialize the mixer bias DAC values.
values.
PRESET
, −2001,
FREQUENCY
DISPLAY
CHANGE TITLE
value to 1600 for band 1.
nonvolatile memory.
,
,
CAL
More 1 of 4
More 2 of 4
SERVICE CAL
EXECUTE TITLE
2-5, RF section troubleshooting is
2-5, complete the rest of this procedure to
Hz
Acceptable
Bias Range
800 to 1900
0 to 200
1000 to 1800
400 to 750
Chapter 2