Agilent Technologies 8590 Series Service Manual page 208

Analyzers assembly-level repair
Table of Contents

Advertisement

Backing Up and Reloading Correction Constants
Reloading the Correction Constants
Reload the A12 step-gain-correction constants
1. Reload the A12 step-gain-correction constants by pressing the
following keys.
PRESET
, −2001,
FREQUENCY
,
CAL
More 1 of 4
,
Service Cal
SET ATTN ERROR
REF LVL OFFSET is displayed in the active-function block above the
prompt ENTER CAL ATTEN ERROR 1.
2. At the prompt, enter the five step-attenuator correction constants
(resolution to 0.01 dB) listed in the Correction Constant
Backup-Data Record, then terminate each entry with either
−dBm
, as appropriate. Typically 1 and 2 dB step errors are negative.
The 4 and 8 dB steps are positive. The 16 dB step is always 0 dB.
Each entry is displayed to the left of the graticule as an amplitude
offset, but only with 0.1 dB resolution. A PRESET occurs after the
16 dB step-attenuator error is entered.
Reload the differential phase calibration constant
This step is for instruments equipped with Option 107 only.
1. Install 85721A Cable TV Measurements Personality.
Refer to your cable TV measurements user's guide for the procedure
to load this personality.
Press
FREQUENCY
Enter −2001,
Hz
Press
CAL, More 1 of 4
Press
Service Cal
Press
Store DP CAL
Enter −7373,
Hz
208
Hz
,
More 2 of 4
,
More 2 of 4
,
Flatness Data
of
+dBm
Chapter 3