Agilent Technologies 8590 Series Service Manual page 600

Analyzers assembly-level repair
Table of Contents

Advertisement

Softkey Descriptions
Softkey Descriptions
INIT FLT 22 GHz Initialize Flatness 22 GHz
D
The current flatness-correction constants are lost when
CAUTION
INIT FLT 22 GHz
R
ELATED
INIT FLT 26.5 GHz Initialize Flatness
26.5 GHz
D
The current flatness-correction constants are lost when
CAUTION
or
600
Note that this function is available for 8592L and
ESCRIPTION
8593E spectrum analyzers only.
Prepares the analyzer memory for the entry of new
flatness correction constants. The −2001 Hz passcode is
required and must be entered before pressing
INIT FLT 22 GHz
• Sets up the start frequency, stop frequency, and step
size of the frequency-response-correction points up
to 22 GHz.
• Initializes A16 memory by entering a default value
of 0 dB into memory for all frequency points.
, or
INIT FLT 26.5 GHz
S
EDIT FLATNESS
OFTKEYS
INIT FLT 26.5 GHz
STORE FLATNESS
Be sure to use
ESCRIPTION
Option 026, 26.5 GHz extended-frequency range, only.
Prepares the analyzer memory for the entry of new
flatness correction constants. The −2001 Hz passcode is
required and must be entered before pressing
INIT FLT 26.5 GHz
• Sets up the start frequency, stop frequency, and step
size of the frequency-response-correction points up
to 26.5 GHz.
• Initializes A16 memory by entering a default value
of 0 dB into memory for all frequency points.
is pressed.
INIT FLT 26.5 GHz
performs the following functions.
is pressed.
for instruments with
INIT FLT 26.5 GHz
performs the following functions.
.
CAL
,
INIT FLT
.
CAL
INIT FLT 22 GHz
Chapter 13