Table 1-7
Adjustments and Tests for Replaced or Repaired Assemblies
Replaced or Repaired
Assembly
A3A14 LO Distribution
Amplifier (LODA)
A3A15 Tracking
Generator
A3 Input Attenuator
A4 First Converter
A5 Second Converter
A6 YIG-Tuned
Oscillator
Chapter 1
Related Adjustments
and Adjustment
Routines
First LO
Distribution Amplifier
for Option 009 or 010
only
CAL FREQ
CAL AMPTD
CAL YTF
First LO
Distribution Amplifier
for Option 009 or 010
CAL FREQ
CAL AMPTD
CAL YTF
BITG Power Level for
Option 010
CAL FREQ
CAL AMPTD
Frequency Response
CAL FREQ
CAL AMPTD
Second Converter
CAL FREQ
CAL AMPTD
CAL FREQ
CAL AMPTD
If You Replace or Repair an Assembly
Related Performance Verification Tests
Frequency Readout and Marker Count Accy
Noise Sidebands
System Related Sidebands
Residual FM
Frequency Span Readout Accuracy
Frequency Response
Other Input Related Spurious Responses
Spurious Response
Residual Responses
Frequency Readout and Marker Count Accy
Noise Sidebands
System Related Sidebands
Residual FM
Frequency Span Readout Accuracy
Frequency Response
Other Input Related Spurious Responses
Spurious Response
Residual Responses
Frequency Response
Noise Sidebands
System Related Sidebands
Frequency Response
Spurious Response
Gain Compression
Displayed Average Noise
Residual Responses
Noise Sidebands
Residual FM
System Related Sidebands
Frequency Response
Spurious Response
Gain Compression
Displayed Average Noise
Residual Responses
Noise Sidebands
Residual FM
System Related Sidebands
Frequency Span Readout Accuracy
Frequency Response
Spurious Response
Displayed Average Noise
Residual Responses
Introduction