Agilent Technologies 8590 Series Service Manual page 639

Analyzers assembly-level repair
Table of Contents

Advertisement

Analyzer Messages
Interpreting Analyzer Messages
Gate card not calibrated (U)(H)
This message can indicate that either the CAL AMPTD routine need to
be performed before the time-gating functions can be used, or that
something was connected to the GATE TRIGGER INPUT connector
during the CAL AMPTD or CAL FREQ &AMPTD routines.
Related Assemblies
A105 gated video board assembly.
Troubleshooting Hints
If your analyzer is equipped with an Option 105, ensure that nothing is
connected to the GATE TRIGGER INPUT connector when the CAL
AMPTD or CAL FREQ &AMPTD routines are performed.
INTERNAL LOCKED (U)
The analyzer internal trace and state registers have been locked. To
unlock the trace or state registers, press SAV LOCK ON OFF so that
OFF is underlined. For remote operation, use PSTATE OFF.
INVALID ACTDEF: _ _ _ (U)
The specified ACTDEF name is not valid. See the ACTDEF
programming command.
INVALID AMPCOR: FREQ (U)
For the AMPCOR command, the frequency data must be entered in
increasing order. See the description for the AMPCOR programming
command for more information.
INVALID BLOCK FORMAT: IF STATEMENT (U)
An invalid block format appeared within the IF statement. See the
description for the IF THEN ELSE ENDIF programming command for
more information.
INVALID CARD (U)
Indicates one of the following conditions: a card reader is not installed,
the memory card is write-protected (check the position of the switch on
the memory card), the memory card is a read-only memory (ROM) card,
or a memory card has not been inserted. This message can also occur if
remote programming commands for the memory card capability are
executed with an 8590L or 8592L that does not have an Option 003.
Chapter 14
639