Agilent Technologies 8590 Series Service Manual page 429

Analyzers assembly-level repair
Table of Contents

Advertisement

IS THE FAILURE IN A SINGLE CHANNEL BANDWIDTH?
Comments
Connector
Problems
Care must be taken
on inserting the
cables into the board
to keep the male
connectors well
aligned with the
female connectors.
Inspect the cable
male connector
before EVERY
insertion. Correct
any tilt or off-center
of the center pin, if
necessary.
Chapter 10
First LO Distribution Amplifier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
TREE 4: EVM Problems
Problems/Reasons
The MCX connectors used on the
gray cables on the option cards
are delicate.
Analyzer Options and Block Diagrams
Diagnostic/Fix Action
Inspect both the female connector
on the board, and the male
connector on the cable.
If the female connector on the board
is not severely crushed, it can often
be cleared using a dental pick. If the
male connector pin appears to be off
center, recenter using fine tweezers.
Replace the board only if the
connector center hole is crushed
beyond repair.