LC Alignment
18.Press the following analyzer keys.
19.Widen all but one of the LC filter poles by shorting A11TP10 and
Use a tool with a nonmetallic body to make the LC dip adjustment.
CAUTION
Shorting components to ground may result in permanent damage to the
bandwidth board assembly.
20.Center the signal on the analyzer display by pressing the following
21.Adjust A13C47 LC dip for a minimum signal amplitude.
22.Move the short from A13TP10 and A13TP11 to A13TP12 and
23.Move the short from A11TP10 and A11TP11 to A13TP10 and
24.Move the short from A11TP12 and A11TP13 to A11TP10 and
LC Centering
Note that the center frequency of the 100 kHz bandwidth is referenced
to the 30 kHz bandwidth. During this procedure it is advisable to
switch to the 30 kHz bandwidth occasionally and recenter it using
PEAK SEARCH
25.Short A11TP10 and A11TP11, A11TP12 and A11TP13, and
Chapter 2
, 100,
BW
kHz
, 5,
SPAN
MHz
A11TP11, A11TP12 and A11TP13, and A13TP10 and A13TP11
using the crystal shorts.
analyzer keys.
PEAK SEARCH
MKR FCTN, MK TRACK ON OFF
A13TP13, then adjust A13C17 LC dip for a minimum signal
amplitude.
A13TP11, then adjust A11C17 LC dip for a minimum signal
amplitude.
A11TP11, then adjust A11C47 LC dip for a minimum signal
amplitude.
MARKER → CF.
,
A13TP10 and A13TP11 using the crystal shorts used in the crystal
alignment section. Press the following analyzer keys.
, 30,
BW
kHz
, 200,
SPAN
kHz
,
MKR FCTN
MK TRACK ON OFF
, 100,
BW
kHz
5. Crystal and LC Bandwidth Filter
(ON)
(OFF)
Making Adjustments