Agilent Technologies 8590 Series Service Manual page 640

Analyzers assembly-level repair
Table of Contents

Advertisement

Analyzer Messages
Interpreting Analyzer Messages
INVALID CARD: BAD MEDIA (U)(H)
Indicates the formatting routine (
) for the memory card
FORMAT CARD
could not be completed. See the description for INVALID CARD above
for more information about the possible causes of this message.
INVALID CARD: DATA ERROR (U)(H)
Indicates the data could not be retrieved from the memory card.
INVALID CARD: DIRECTORY (U)
Indicates the memory card has not been formatted.
INVALID CARD: NO CARD (U)
Indicates a memory card has not been inserted.
INVALID CARD: TYPE (U)
Indicates one of the following conditions: a card reader is not installed,
the memory card is write-protected (check the position of the switch on
the memory card), the memory card is a read-only memory (ROM) card,
or a memory card has not been inserted. This message can also occur if
remote programming commands for the memory card capability are
executed with an 8590L, 8592L, or 8594L that does not have an
Option 003.
INVALID CHECKSUM: USTATE (U)
The user-defined state does not follow the expected format.
INVALID COMPARE OPERATOR (U)
An IF/THEN or REPEAT/UNTIL routine is improperly constructed.
Specifically, the IF or UNTIL operands are incorrect.
INVALID DET: FM or TV option only (U)
Indicates that the selected detector cannot be used until the
appropriate option is installed in the analyzer.
INVALID ENTER FORMAT (U)
The enter format is not valid. See the appropriate programming
command description to determine the correct format.