Agilent Technologies 8590 Series Service Manual page 194

Analyzers assembly-level repair
Table of Contents

Advertisement

Making Adjustments: If 3335A Source Not Available
15a. Frequency Response of the 8590L, 8591C, and 8591E
4. On the analyzer, press the following keys.
PRESET
CAL, More 1 of 4
CORRECT ON OFF
, 300,
FREQUENCY
CF STEP AUTO MAN
, 12,
SPAN
MHz
75 Ω input: press
, 10,
AMPLITUDE
SCALE LOG LIN
, 1,
BW
MHz
PEAK SEARCH
,
MKR FCTN
MK TRACK ON OFF
5. Adjust the synthesized signal generator POWER LEVEL for a
MKR-TRK amplitude reading of −14 dBm ±0.1 dB.
6. Press RATIO mode on the measuring receiver.
7. Press the following analyzer keys.
, 4,
FREQUENCY
8. Set the synthesized signal generator CW to 4 MHz.
9. Adjust the synthesized signal generator POWER LEVEL for an
analyzer MKR-TRK amplitude reading of −14 dBm ±0.1 dB.
10.Record the power ratio in Column 2 of
4 MHz.
11.Press the following analyzer keys.
, 41,
FREQUENCY
12.Set the synthesized signal generator CW to 41 MHz.
13.Adjust the synthesized signal generator POWER LEVEL for an
analyzer MKR-TRK amplitude reading of −14 dBm ±0.1 dB.
14.Record the power ratio displayed on the measuring receiver in
Column 2 of
Table 2a-5
15.On the analyzer, press
the remaining frequencies listed in Column 1 of
new frequency repeat
Cal Factor into the measuring receiver as indicated in Column 3 of
Table
2a-5.
194
(OFF)
MHz
, 37,
MHz
,
AMPLITUDE
More 1 of 2
−dBm
(LOG), 1,
dB
(ON)
MHz
MHz
for 41 MHz.
and
FREQUENCY
step 13
to
step
15, entering the power sensor
,
,
AMPTD UNITS
dBm
Table 2a-5
for
(step up), to step through
Table
2a-5. At each
Chapter
2a