15.Enter the Corrected Flatness ERROR from
for the frequency displayed using the DATA keys. Terminate the
entry with the
+dBm
After the new data is entered, the analyzer will automatically jump
to the next frequency correction point. Enter the next Corrected
Flatness ERROR from the table.
To confirm the entries, press
interest. Re-check the displayed error against the Corrected
Flatness ERROR from column 3.
16.Press More, EXIT after all corrections are complete.
17.Perform the frequency response verification test from the calibration
guide for your instrument. Only checking Band 0 is necessary.
Chapter 2
28. Correcting for Absolute Amplitude Accuracy for Option 051
.
(step down key) to the frequency of
Making Adjustments
Table 2-28
or
Table 2-29