Making Adjustments
11. CAL MXR Adjustment Routine
Procedure
1. Perform the CAL YTF routine as indicated in the "CAL YTF
Adjustment Routine" in this chapter.
2. Connect the 100 MHz COMB OUT to INPUT 50 Ω using the CAL
YTF cable (Type N to SMA cable). Refer to
Model 8595E only: Connect the CAL OUT to the INPUT 50 Ω using a
BNC cable.
Figure 2-16
CAL MXR Adjustment Setup
3. Press the following analyzer keys.
Note that the frequency of −2001
SERVICE CAL
The
"CAL: DONE" will be displayed when the routine has finished. The
internal adjustment data will be stored in working RAM.
4. Press
78
PRESET
, −2001,
FREQUENCY
,
CAL, More 1 of 4
More 2 of 4
,
SERVICE CAL
CAL MXR
routines.
routine will take a few minutes to run. The message
CAL MXR
to store the correction data in nonvolatile memory.
CAL STORE
Figure
Hz
is necessary to access the
Hz
2-16.
Chapter 2