Agilent Technologies 8590 Series Service Manual page 84

Analyzers assembly-level repair
Table of Contents

Advertisement

Making Adjustments
12. Second Converter
21.Press
marker delta amplitude reads 0 dB and is on the right-hand side of
the response. Read the marker delta frequency; it should be 15 MHz
±2.5 MHz. See
22.The maximum ripple should be less than 1 dB for a bandwidth of
6 MHz.
Figure 2-21
Second Converter Bandpass Ripple Measurement
84
and rotate the knob clockwise so that the
MARKER MODE,
Figure
2-21.