Troubleshooting the IF Section
IF Power-Level Measurement
Table 5-4
Calibration-Attenuator Control Lines
Reference
Calibratio
Level
n
(dBm)
Attenuato
r
0
10
−1
9
−2
8
−3
7
−4
6
−5
5
−6
4
−7
3
−8
2
−9
1
H = >2.5 V (TTL High)
L = <0.8 V (TTL Low)
264
IFA1
IFA2
(1 dB)
(2 dB)
A12J2-4
A12J2-3
L
H
H
L
L
L
H
H
L
H
H
L
L
L
H
H
L
H
H
L
IFA3
IFA4
(4 dB)
(8 dB)
A12J2-2
A12J2-1
L
H
L
H
L
H
H
L
H
L
H
L
H
L
L
L
L
L
L
L
IFA5
(16 dB)
A12J2-11
L
L
L
L
L
L
L
L
L
L