55.Press the following analyzer keys.
56.Step through the remaining frequencies listed in
Entering Flatness Correction Data
57.Enter the pass code by pressing the following analyzer keys.
58.To access the flatness correction menu, press the following analyzer
The next step will erase all current flatness correction. This step is to be
CAUTION
performed only if the A16 processor/video assembly has been repaired
or replace.
59.Initialize the area of memory where the flatness correction data is
60.Enter the flatness corrections, by pressing
61.The frequency of the first data point, 12.00 MHz, will be displayed in
Chapter 2
16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
,
MKR
MARKERS OFF
FREQUENCY, ⇑
(Option 026:
Table
2-13). At each new frequency, repeat
step 55
and enter the appropriate power sensor cal factor into the
measuring receiver as listed in Column 2 of
Table
2-13).
Adjustment for the 8592L and 8593E spectrum analyzers is
complete. Continue with the next step.
PRESET
, −2001,
FREQUENCY
keys.
,
CAL, More 1 of 4
More 2 of 4
SERVICE CAL
FLATNESS DATA
stored, by pressing the following keys.
For Option 026, 027 only: Use
INIT FLT 22 GHz
PRESET
, −2001,
FREQUENCY
,
CAL, More 1 of 4
More 2 of 4
SERVICE CAL
FLATNESS DATA
the active function block of the analyzer display.
Table 2-12
Hz
Hz
EDIT FLATNESS
Making Adjustments
Table 2-12
step 51
to
(Option 026:
INIT FLT 26.5 GHz
.
113