4. Place the instrument on the work surface so that it is resting on its
5. Position the display assembly part of the way into the front of the
6. Begin to push W2 and W3 out through the slot (4) in the instrument
A corner of the A8 Power Supply partially blocks the slot (4) in the
CAUTION
instrument chassis. Make sure that the display's wiring harness and
connectors do not jam between the power supply and the chassis.
7. Replace the display in the instrument while carefully pulling W2
8. Hold the display as far in as possible and replace the two screws (5)
9. Connect the W51 wire harness to the power supply and, on 8592L
10.Place the instrument on the work surface with the bottom side
11.Remove the tape protecting the plastic clip on W3 (3).
12.Connect W2 (2) to A16J8 on the A16 Processor/Video Board
13.Replace the three screws (1) that secure the display to the chassis.
14.Replace the A1 Front-Frame assembly. Refer to the
15.Replace the A17 Memory Card Reader. Refer to the
16.Replace the instrument cover assembly. Refer to the
Chapter 7
right side.
instrument.
chassis. See
Figure
7-2.
and W3 through the slot (4) in the instrument chassis.
that secure the display to the left side of the instrument chassis.
and 8593E spectrum analyzers, to the A7 Analog Control assembly.
See
Figure
7-8.
facing up.
assembly.
Assembly."
replacement procedure.
Card Reader Assembly."
Cover."
replacement procedure.
replacement procedure.
Replacing Major Assemblies
A2 Display
"A1 Front-Frame
"A17 Memory
"Instrument
307