Agilent Technologies 8590 Series Service Manual page 66

Analyzers assembly-level repair
Table of Contents

Advertisement

Making Adjustments
7. Log and Linear Amplifier
Figure 2-12
Log and Linear Amplifier Adjustment Location
12.Set the synthesizer amplitude 60 dB below that recorded in step 8 by
pressing
Adjust A14R10 OFFSET for the DMM reading of 250 mV ±1 mV.
13.Repeat step 10 and step 11 until no further adjustment is necessary.
14.Set the synthesizer amplitude 30 dB below that recorded in step 8
and adjust the A14R23 SLOPE for a DMM reading of 625 mV ±1 mV.
15.Set the synthesizer amplitude to the level recorded in step 8 and
adjust the A14R69 −30 dB for a DMM reading of 1000 mV ±1 mV.
16.Repeat step 13 and step 14 until no further adjustment is necessary.
17.Set the synthesizer amplitude 10 dB below that recorded in step 8
and adjust the A14R23 SLOPE for a DMM reading of 875 mV ±1 mV.
18.Set the synthesizer amplitude to the level recorded in step 8 and
adjust the A14R39 −10 dB for a DMM reading of 1000 mV ±1 mV.
19.Repeat step 16 and step 17 until no further adjustment is necessary.
20.Repeat step 10 to step 18 until the limits in
66
and then pressing
AMPLITUDE
(step-down key) six times.
Table 2-3
are met.
Chapter 2