7. Press the following keys to allow entry of the default mixer bias DAC
value for band 2.
DISPLAY
CHANGE TITLE
CLEAR
8. Enter CALMXRDATA 20,2; as a title entry to set the mixer bias value
to 20 for band 2.
Press the following keys to store the mixer bias value for band 2 in
nonvolatile memory.
CAL, More 1 of 4
SERVICE CAL
EXECUTE TITLE
9. Press the following keys to allow entry of the default mixer bias DAC
value for band 3.
DISPLAY
CHANGE TITLE
CLEAR
10.Enter CALMXRDATA 1300,3; as a title entry to set the mixer bias
value to 1300 for band 3.
11.Press the following keys to store the mixer bias value for band 3 in
nonvolatile memory.
CAL, More 1 of 4
SERVICE CAL
EXECUTE TITLE
12.Press the following keys to allow entry of the default mixer bias DAC
value for band 4.
DISPLAY
CHANGE TITLE
CLEAR
13.Enter CALMXRDATA 520,4; as a title entry to set the mixer bias
value to 520 for band 4.
Chapter 2
,
More 2 of 4
,
More 2 of 4
Making Adjustments
10. CAL YTF Adjustment Routine
75