Agilent Technologies 8590 Series Service Manual page 617

Analyzers assembly-level repair
Table of Contents

Advertisement

Before removing the A16 processor/video assembly, check all assemblies
that are identified in
1. Turn the instrument power off and remove each assembly, one at a
time.
2. Turn the instrument on.
• If the ADC ground fault remains, the instrument preset at
power-up displays the error message.
• If the error message goes away, the last assembly removed is the
cause of the problem.
• If the error message remains after the removal of all related
assemblies, suspect the A16 processor/video assembly.
An ADC ground failure can also be checked manually, using the service
diagnostic. Press
CAL
.
REF DETECTOR
Chapter 14
Table
14-1.
,
,