A17 Memory Card Reader Assembly

This procedure applies to:
Removal
1. Remove the instrument cover assembly. Refer to the
2. Place the analyzer on the work surface with the bottom side facing
3. Disconnect the W14 ribbon cable (1) from the A17 Memory Card
4. Remove the jumper (5) between the A17 Memory Card Reader and
5. Remove the two screws (2) that secure the A17 assembly to the
When removing the A17 assembly, be sure to slide the assembly
CAUTION
straight backward to prevent damage to the two alignment tabs on the
front of the A17 support bracket.
6. Remove the A17 assembly (3), gently sliding the assembly towards
7. Remove the four screws (4) that secure the A17 assembly to the
Chapter 7
8590 E-Series and L-Series spectrum analyzers
8591C cable TV analyzers
8594Q QAM analyzers
Cover."
removal procedure.
up.
Reader assembly. See
LS1. See
Figure
7-16.
chassis.
the rear of the analyzer and up.
memory-card base and remove the A17 assembly.
A17 Memory Card Reader Assembly
Figure
7-16.
Replacing Major Assemblies
"Instrument
339