Agilent Technologies 8590 Series Service Manual page 29

Analyzers assembly-level repair
Table of Contents

Advertisement

Table 1-7
Adjustments and Tests for Replaced or Repaired Assemblies
Replaced or Repaired
Assembly
A11 Bandwidth
A12 Amplitude Control CAL FREQ
A13 Bandwidth
A14 Log Amplifier
A15 Motherboard
A16 Processor/Video
A16A1 Memory
† These adjustments are not necessary if valid correction constants are recovered from the
defective A16 processor/video assembly. See the A16 processor/video board assembly replacement
procedure in this service guide.
Chapter 1
Related Adjustments
and Adjustment
Routines
Crystal and IC
Bandwidth Filter CAL
FREQ
CAL AMPTD
CAL AMPTD
Crystal and IC
Bandwidth Filter CAL
FREQ
CAL AMPTD
Log and Linear
Amplifier
CAL FREQ
CAL AMPTD
CAL FREQ
CAL AMPTD
10 MHz Reference
CAL FREQ
CAL AMPTD
CAL YTF (8592L,
8593E, 8595E, 8596E
only)
CAL MXR † (8592L,
8593E, 8595E, 8596E
only)
Cal Attenuator Error †
Frequency Response †
Time and Date
If You Replace or Repair an Assembly
Related Performance Verification Tests
Resolution Bandwidth Accuracy
Absolute Amplitude Calibration and
Bandwidth Switching Uncertainties
Scale Fidelity
Displayed Average Noise
Resolution Bandwidth Accuracy
Absolute Amplitude Calibration and
Bandwidth Switching Uncertainties
Scale Fidelity
Reference Level Accuracy
Complete Operation Verification
Reference Level Accuracy
Frequency Response
Introduction
29

Advertisement

Table of Contents
loading