Making Adjustments
13. Third Converter and Second IF Bandpass
28.Subtract the Mismatch Error
Insertion Loss
For example, if the Mismatch Error is +0.3 dB and the uncorrected
Insertion Loss is −10.2 dB, subtracting the mismatch error to the
insertion loss gives a corrected reading of −10.5 dB.
300 MHz Calibrator Amplitude Adjustment
29.Connect the equipment as shown in
should be positioned so that the setup of the adapters, LPF and
attenuator do not bind. It may be necessary to support the center of
gravity of the devices.
Figure 2-24
300 MHz Calibrator Amplitude Accuracy Test Setup
30.On the power meter, press the dBm mode key. Record the power
meter reading in dBm.
31.Subtract the Corrected Insertion Loss
meter reading
OUT should be −20 dBm ±0.4 dB.
90
(step
22). This is the corrected insertion loss.
Corrected Insertion Loss____________________dB
Power Meter Reading____________________dBm
(step
25) and record as the CAL OUT power. The CAL
CAL OUT Power = Power Meter Reading − Corrected Insert Loss
(step
22) from the Uncorrected
Figure
2-24. The analyzer
(step
23) from the power
Chapter 2