6a. 8590L, 8591C, and 8591E

6a. 8590L, 8591C, and 8591E
This section describes the RF section troubleshooting process for the 8590L
and 8591E spectrum analyzers and 8591C cable TV analyzers.
Making RF Power-Level Measurements
The power level ranges listed for measurements A through G in
Table 6-1 apply after performing the following steps:
1. Connect the CAL OUT to the RF INPUT using the CAL cable.
2. Perform the frequency and amplitude calibration routines.
3. Press
4. Press the following instrument keys:
Table 6-1
Power Levels at Measurement Points
Measurement Point
A
B
C
D
E
F
G
* A frequency-selective measuring device, such as another analyzer, is recommended for making these
measurements. Broadband measuring devices, such as power meters, give erroneous results due to the presence
of other, higher-amplitude signals.
284
Ensure that the CAL OUT amplitude is within specification. (Refer to
the calibrator amplitude performance test in the calibration guide.)
, then wait for the analyzer to complete the preset routine.
PRESET
, 300,
FREQUENCY
MHz
, 0,
SPAN
Hz
Measurement
Frequency
300 MHz
2.1214 GHz
2.1214 GHz
321.4 MHz
2.4214 GHz
2.4214 GHz
1.8 GHz
Power Level Range *
(dBm)
50 Ω
−29 to −31
−39 to −43
−39 to −44
−42 to −46
+6 to +12
−10 to −20
−17 to −23
75 Ω
−35 to −37
−46 to −49
−46 to −50
−51 to −55
Same as 50 Ω
Same as 50 Ω
Same as 50 Ω
Chapter 6